V1.2 01/11/10 Time Electronics Ltd Botany Industrial Estate, Tonbridge, Kent, TN9 1RH Tel: +44(0)1732 355993 Fax: +44(0)1732 770312 Email: mail@T
10 1017 Technical Manual 10 5. Guarantee & Servicing Guarantee Period This unit is guaranteed against defects in materials and workmanship for a
2 1017 Technical Manual 2 Contents 1. Introduc tion ...
3 1017 Technical Manual 3 1. Introduction 1.1. General Description The 1017 is a high performance portable calibrator for use in the field or labor
4 1017 Technical Manual 4 1.2. Specifications Outputs: 0 – 100V in 5 Ranges 0 – 100mA in 1 Range 0 – 10kΩ in 1 Range Voltage
5 1017 Technical Manual 5 Dimensions & Weight Dimensions – 290mm x 250mm x 110mm Weight – 2.4Kg Operating Temperature 0°C to +50°C max. For
6 1017 Technical Manual 6 2. Front Panel Controls 1 – Output Terminals. 2 – Case Terminal. 3 – Reverse / Off / Normal Switch, Selects the out
7 1017 Technical Manual 7 3. Operation 3.1. Preparing for Use Before using the 1017, you must first check to see if the battery level indicator shows
8 1017 Technical Manual 8 4. Calibration 4.1. Calibrating the 1017 The 1017 can be calibrated with a high accuracy D.M.M. with D.C. voltage, current
9 1017 Technical Manual 9 4.3. Calibrating the Full Scales After checking the zero settings you may then calibrate the full scales. Set the digit swi
Comments to this Manuals